Washer-Disinfector Temperature — MFDS Part 1 MCQ
Instant feedback + full explanation. One question, done properly.
Educational content. Not a substitute for clinical judgement or local policy.
Reveal the answer and explanation
Correct answer: B — 80°C for 10 minutes or 90°C for 1 minute
Explanation lettering: C = shown as A · A = shown as B · B = shown as C
A is correct. UK dental decontamination guidance gives 80°C for 10 minutes or 90°C for 1 minute as examples of the thermal-disinfection holding phase in a washer-disinfector cycle. The higher temperature permits the shorter holding time. Option B contains the key error in the original item: 80°C for 1 minute is not the cited exposure. Options C and E are cooler and/or shorter than the specified examples, while option D gives much shorter exposures. Thermal disinfection is one stage of automated instrument reprocessing; after cleaning, disinfection and inspection, reusable instruments still require the appropriate validated steam-sterilisation process.
Reference: Scottish Health Technical Memorandum 01-05: Management, equipment, and process of the decontamination of dental instruments in a Local Decontamination Unit in NHSScotland, Part C, 2024, as identified and endorsed by SDCEP. https://www.sdcep.org.uk/published-guidance/decontamination/