Following microbiological advice, a biological indicator is being used during an exceptional investigation of a steam steriliser. Which microorganism should the indicator contain?

Educational content. Not a substitute for clinical judgement or local policy.

Reveal the answer and explanation

Correct answer: CSpores of Geobacillus stearothermophilus

The correct answer is C. Geobacillus stearothermophilus spores are the specified biological challenge for steam sterilisation because of their resistance to moist heat. Biological indicators assess whether test microorganisms survive the process, but UK guidance states that they should not be used for routine monitoring of steam sterilisers; exceptional use should follow advice from a microbiologist with decontamination expertise. Routine assurance instead relies on validated cycle parameters, automatic control tests and the applicable steam-penetration tests. Bacillus atrophaeus is the most plausible distractor because its spores are used principally to challenge dry-heat and certain chemical processes, not wet-heat steam sterilisation. The other listed spore-forming organisms are not the specified reference biological indicators for this purpose.

Reference: NHS England, Health Technical Memorandum 01-01: Management and decontamination of surgical instruments, Part C: Steam sterilization, sections 2.100–2.103, 2016. https://www.england.nhs.uk/wp-content/uploads/2021/05/HTM0101PartC.pdf