Handpiece Sterilisation — MFDS Part 1 MCQ
Instant feedback + full explanation. One question, done properly.
Educational content. Not a substitute for clinical judgement or local policy.
Reveal the answer and explanation
Correct answer: B — Steam sterilisation in a validated compatible sterilizer
Explanation lettering: C = shown as A · A = shown as B · B = shown as C
A is correct. A reusable, steam-compatible dental handpiece should be cleaned and lubricated as specified by its manufacturer, then sterilised using a validated compatible steam sterilizer. UK dental decontamination guidance identifies saturated steam under pressure as the preferred method for sterilising most reusable clinical instruments, with the sterilizer cycle and load selected according to manufacturer compatibility. A fixed cycle such as 134–137°C for three minutes should not be asserted for every handpiece, because compatibility and validated cycles vary. Glutaraldehyde, alcohol wipes and ultraviolet exposure do not provide validated sterilisation of the handpiece, including its internal components. Dry heat is not the standard UK method for routine processing of steam-compatible dental handpieces and may not be compatible with the device.
Reference: NHS England, Health Technical Memorandum 01-05: Decontamination in primary care dental practices, section 4 (Sterilization), 2013; webpage updated 26 January 2024. https://www.england.nhs.uk/publication/decontamination-in-primary-care-dental-practices-htm-01-05/