Arterial transducer levelling — FRCA Primary MCQ
Instant feedback + full explanation. One question, done properly.
Educational content. Not a substitute for clinical judgement or local policy.
Reveal the answer and explanation
Correct answer: A — Subtract approximately 15 mmHg from the displayed pressure
A pressure transducer below the reference level over-reads because the fluid column adds hydrostatic pressure; 20 cmH2O is about 15 mmHg. The tempting error is to add pressure because the patient appears hypotensive, but the displayed pressure is already falsely high. A normal waveform does not correct for an incorrect zero reference height. The transducer should be levelled to the phlebostatic axis.
Reference: RCoA Primary FRCA Syllabus 2026; Davis and Kenny Basic Physics and Measurement in Anaesthesia