Diathermy and pacemaker safety — FRCA Primary MCQ
Instant feedback + full explanation. One question, done properly.
Educational content. Not a substitute for clinical judgement or local policy.
Reveal the answer and explanation
Correct answer: A — Tachytherapy stops; programmed bradycardia pacing remains unchanged
The best answer is “Tachytherapy stops; programmed bradycardia pacing remains unchanged”. A magnet over a transvenous ICD generally inhibits tachyarrhythmia detection and therapy but does not alter its bradycardia pacing mode. A pacemaker-dependent patient exposed to electromagnetic interference may therefore still have pacing inhibited and usually needs an explicit reprogramming plan, monitoring and restoration of therapies afterwards.
Reference: British Heart Rhythm Society, Guidelines for the peri-operative management of people with cardiac implantable electronic devices: https://doi.org/10.1111/anae.15728