The corrected flow time (FTc) measured by oesophageal Doppler is a measure of:

Educational content. Not a substitute for clinical judgement or local policy.

Reveal the answer and explanation

Correct answer: ALeft ventricular preload (filling status)

FTc is the systolic flow time corrected for heart rate (analogous to corrected QT). Normal FTc is approximately 330-360 ms. FTc <330 ms suggests hypovolaemia (shortened filling time). FTc >360 ms suggests vasodilation or fluid overload. It is used as a goal-directed fluid therapy parameter during surgery.

Reference: Al-Shaikh Stacey 2019 Essentials of Anaesthetic Equipment