Hypoxic Guard O2 N2O Link — FRCA Primary MCQ
Instant feedback + full explanation. One question, done properly.
Educational content. Not a substitute for clinical judgement or local policy.
Reveal the answer and explanation
Correct answer: E — Delivery of less than 21% oxygen (or 25% depending on manufacturer) by mechanically or pneumatically linking the O2 and N2O flow controls
The hypoxic guard (e.g. Link-25 / Ratio Mixer / S-ORC) mechanically or pneumatically couples the O2 and N2O flowmeter controls so that the minimum O2 concentration delivered is 21-25%. It does NOT protect against hypoxia from other causes (wrong pipeline / cylinder depletion / circuit disconnection / air dilution). Some newer machines use electronic proportioning systems.
Reference: Al-Shaikh Stacey 2019 Essentials of Anaesthetic Equipment