ModeratePhysics & Clinical MeasurementNIBP Oscillometry MAP DirectFRCA Primary

Regarding non-invasive blood pressure (NIBP) measurement by oscillometry the MAP is:

Educational content. Not a substitute for clinical judgement or local policy.

Reveal the answer and explanation

Correct answer: CDirectly measured as the point of maximum oscillation amplitude

Oscillometric NIBP: the MAP is the directly measured value (corresponding to the point of maximum oscillation amplitude as the cuff deflates). Systolic and diastolic pressures are calculated from the MAP using proprietary algorithms which vary between manufacturers. This is why MAP is the most accurate oscillometric measurement and systolic/diastolic values may vary between devices.

Reference: Al-Shaikh Stacey 2019 Essentials of Anaesthetic Equipment