Splitting Ratio Vaporiser — FRCA Primary MCQ

Instant feedback + full explanation. One question, done properly.

ModeratePhysics & Clinical MeasurementSplitting Ratio VaporiserFRCA Primary

Regarding vaporiser output the splitting ratio refers to:

Educational content. Not a substitute for clinical judgement or local policy.

Reveal the answer and explanation

Correct answer: BThe ratio of gas that flows through the vaporising chamber to gas that bypasses it

In a variable bypass (plenum) vaporiser the total fresh gas flow is split: a small proportion flows through the vaporising chamber (becoming saturated with vapour) and the majority bypasses it. The ratio is adjusted by the concentration dial. The splitting ratio determines the final output concentration. Temperature compensation adjusts this ratio to maintain constant output.

Reference: Al-Shaikh Stacey 2019 Essentials of Anaesthetic Equipment