Regarding electrical safety in theatres the isolated (floating) power supply reduces the risk of electrocution by:

Educational content. Not a substitute for clinical judgement or local policy.

Reveal the answer and explanation

Correct answer: EProviding no earth reference so that a single fault does not complete a circuit through the patient to earth

An isolated power supply (transformer with no earth reference on the secondary winding) means that a single fault (patient touching one live wire) does not complete a circuit to earth — current cannot flow through the patient. A line isolation monitor detects current leakage to earth and alarms if it exceeds a safe limit (usually 2-5 mA).

Reference: Al-Shaikh Stacey 2019 Essentials of Anaesthetic Equipment