O2 Concentrator PSA Zeolite — FRCA Primary MCQ
Instant feedback + full explanation. One question, done properly.
Educational content. Not a substitute for clinical judgement or local policy.
Reveal the answer and explanation
Correct answer: C — Pressure swing adsorption using zeolite molecular sieves
Oxygen concentrators use pressure swing adsorption (PSA). Compressed air is passed through zeolite molecular sieve columns that preferentially adsorb nitrogen. The remaining gas is >90% oxygen. Two columns alternate: one adsorbs N2 while the other is regenerated by depressurisation. They provide FiO2 up to 95% at flow rates of 1-5 L/min.
Reference: Al-Shaikh Stacey 2019 Essentials of Anaesthetic Equipment