HardPhysics & Clinical MeasurementBain Circuit Mapleson DFRCA Primary

During controlled ventilation with a Bain circuit, inspired carbon dioxide rises abruptly despite high fresh-gas flow and a patent expiratory valve. Which fault is most likely?

Educational content. Not a substitute for clinical judgement or local policy.

Reveal the answer and explanation

Correct answer: BDisconnection of the inner fresh-gas limb

The best answer is “Disconnection of the inner fresh-gas limb”. The Bain is a coaxial Mapleson D circuit. If the inner fresh-gas limb disconnects, fresh gas enters at the machine end and the outer limb effectively becomes apparatus dead space, causing substantial rebreathing. There is no carbon-dioxide absorber in this circuit, and a high nominal flow cannot compensate reliably for the disconnection.

Reference: Kaul and Mittal, Mapleson's breathing systems: https://pmc.ncbi.nlm.nih.gov/articles/PMC3821268/