Bain Circuit Mapleson D — FRCA Primary MCQ
Instant feedback + full explanation. One question, done properly.
Educational content. Not a substitute for clinical judgement or local policy.
Reveal the answer and explanation
Correct answer: B — Disconnection of the inner fresh-gas limb
The best answer is “Disconnection of the inner fresh-gas limb”. The Bain is a coaxial Mapleson D circuit. If the inner fresh-gas limb disconnects, fresh gas enters at the machine end and the outer limb effectively becomes apparatus dead space, causing substantial rebreathing. There is no carbon-dioxide absorber in this circuit, and a high nominal flow cannot compensate reliably for the disconnection.
Reference: Kaul and Mittal, Mapleson's breathing systems: https://pmc.ncbi.nlm.nih.gov/articles/PMC3821268/