O2 Flush Flow Rate — FRCA Primary MCQ

Instant feedback + full explanation. One question, done properly.

ModeratePhysics & Clinical MeasurementO2 Flush Flow RateFRCA Primary

The oxygen flush valve delivers oxygen at approximately:

Educational content. Not a substitute for clinical judgement or local policy.

Reveal the answer and explanation

Correct answer: E35-75 L/min

The oxygen flush delivers 100% oxygen at 35-75 L/min bypassing vaporisers and flowmeters at pipeline pressure (~400 kPa). Use with caution during spontaneous ventilation with a closed circuit as it can generate dangerously high airway pressures.

Reference: Al-Shaikh Stacey 2019 Essentials of Anaesthetic Equipment