The fuel cell oxygen analyser requires periodic replacement because:

Educational content. Not a substitute for clinical judgement or local policy.

Reveal the answer and explanation

Correct answer: EThe lead anode is consumed over time

The galvanic fuel cell generates its own EMF without external power. The lead anode is oxidised over time requiring cell replacement every 6-12 months. Its response time (~20-30 seconds) is too slow for breath-by-breath analysis. The Clark electrode is polarographic and requires an external voltage.

Reference: Al-Shaikh Stacey 2019 Essentials of Anaesthetic Equipment