Vaporiser Output at Altitude — FRCA Primary MCQ
Instant feedback + full explanation. One question, done properly.
Educational content. Not a substitute for clinical judgement or local policy.
Reveal the answer and explanation
Correct answer: D — Approximately 3.3% isoflurane, with a partial pressure of approximately 2.0 kPa
Explanation lettering: E = shown as A · A = shown as D · D = shown as E
Option A is correct. A conventional plenum variable-bypass vaporiser has an approximately fixed splitting ratio at a given dial setting. Because the agent's saturated vapour pressure is principally temperature-dependent, the vaporiser therefore delivers approximately the same anaesthetic partial pressure despite reduced ambient pressure. At sea level, 2.0% of 100 kPa is 2.0 kPa. At 60 kPa, that partial pressure represents 2.0/60 = 3.3% by volume. Option B incorrectly assumes that the dial continues to specify the actual volume percentage. Option C applies the fall in barometric pressure twice. Options D and E give incorrect combinations of concentration and partial pressure. This relationship should not be generalised without qualification to every vaporiser design.
Reference: Royal College of Anaesthetists. Guide to the FRCA Examination: The Primary, Fourth Edition. SBA question 82 and answer commentary, 2013. https://www.rcoa.ac.uk/media/49221