Pressure Monitoring Compliance — FRCA Primary MCQ
Instant feedback + full explanation. One question, done properly.
Educational content. Not a substitute for clinical judgement or local policy.
Reveal the answer and explanation
Correct answer: B — Increasing the length of otherwise identical connecting tubing
Explanation lettering: B = shown as A · D = shown as B · E = shown as C · C = shown as D · A = shown as E
D is correct. Compliance is the volume change produced per unit pressure change. Lengthening otherwise identical tubing adds both fluid volume and distensible tubing wall, so a given pressure change produces a larger total volume displacement. Increased compliance reduces the natural frequency and therefore impairs faithful transmission of rapidly changing pressure waveforms. Changing transducer height alters the hydrostatic pressure offset but not intrinsic compliance. Narrower tubing contains less fluid and has less distensible wall volume, although it also increases resistance. A stiffer diaphragm undergoes less displacement for a given pressure and therefore has lower compliance; reducing diaphragm diameter has the same general effect. The original assertion that increased compliance necessarily causes greater damping was removed because damping ratio also depends on system resistance and inertance.
Reference: Pettersson GB, Kronogård CO, Rådmark T. Studies of the dynamic response of esophageal manometry systems. Scandinavian Journal of Gastroenterology. 1986;21(3):305–312. https://www.rcoa.ac.uk/media/49221