Which description correctly identifies the electrode configuration and operating principle of the Clark polarographic oxygen electrode used in blood-gas analysis?

Educational content. Not a substitute for clinical judgement or local policy.

Reveal the answer and explanation

Correct answer: EPlatinum cathode; silver/silver chloride anode; oxygen is reduced and the current is proportional to PO2

Explanation lettering: E = shown as C · C = shown as D · D = shown as E

D is correct. The Clark electrode is an amperometric, polarographic sensor containing a platinum cathode and silver/silver chloride anode in electrolyte, separated from the sample by an oxygen-permeable membrane. An external polarising potential of approximately 0.6 V is applied. Oxygen diffuses through the membrane and is reduced at the cathode: O2 + 2H2O + 4e− → 4OH−. The diffusion-limited current is proportional to PO2. Option B describes the potentiometric Severinghaus CO2 electrode. Option E describes the principle and typical metals of a galvanic fuel cell, which generates its own electromotive force. Options A and C incorrectly reverse the electrode roles or describe oxygen as undergoing oxidation; reduction occurs at the cathode.