Fick's Law of Diffusion — FRCA Primary MCQ
Instant feedback + full explanation. One question, done properly.
Educational content. Not a substitute for clinical judgement or local policy.
Reveal the answer and explanation
Correct answer: B — Remove air bubbles and shorten the tubing
The best answer is “Remove air bubbles and shorten the tubing”. Air bubbles, compliant components, long tubing, clots and kinks lower natural frequency and increase damping. Removing bubbles and unnecessary tubing therefore improves transmission. Lowering the transducer creates a hydrostatic offset, while excessive pressure-bag inflation does not correct the dynamic response and risks equipment failure.
Reference: Saugel et al., How to measure blood pressure using an arterial catheter: a systematic 5-step approach: https://pmc.ncbi.nlm.nih.gov/articles/PMC7183114/