Bain System FGF Requirement — FRCA Final MCQ

Instant feedback + full explanation. One question, done properly.

ModerateAnaesthetic Equipment & SafetyBain System FGF RequirementFRCA Final

A Bain (coaxial Mapleson D) breathing system is being used for spontaneous ventilation. What fresh gas flow rate is required to prevent rebreathing?

Educational content. Not a substitute for clinical judgement or local policy.

Reveal the answer and explanation

Correct answer: D2-3 times minute ventilation

Mapleson D (Bain): for spontaneous ventilation requires 2-3× minute ventilation FGF to prevent rebreathing (approximately 150-200 mL/kg/min). For controlled ventilation: 70-80 mL/kg/min is sufficient. The Mapleson D is most efficient for controlled ventilation.

Reference: RCOA – 2020 – Primary curriculum: Breathing systems