Machine Leak Common Sources — FRCA Final MCQ
Instant feedback + full explanation. One question, done properly.
Educational content. Not a substitute for clinical judgement or local policy.
Reveal the answer and explanation
Correct answer: A — Loose vaporiser filler cap
Loose vaporiser filler cap = most common leak source. Even with vaporiser off, filler port is part of sealed pathway. Other: loose circuit connections, poorly seated absorber, cracked O-rings. Leak >150 mL/min at 30 cmH2O = fail.
Reference: Association of Anaesthetists – 2023 – Checking anaesthetic equipment