PA Catheter Over-Wedging — FRCA Final MCQ
Instant feedback + full explanation. One question, done properly.
Educational content. Not a substitute for clinical judgement or local policy.
Reveal the answer and explanation
Correct answer: D — Catheter has migrated distally causing persistent wedging (over-wedging)
A dampened PA waveform with diastolic pressure matching PCWP suggests the catheter has migrated distally and become persistently wedged in a smaller PA branch (over-wedging). This is dangerous: the balloon is not inflated but the catheter tip is occluding a vessel, risking pulmonary infarction and PA rupture. Management: immediately withdraw the catheter 2-3 cm (with balloon deflated) until a normal PA waveform returns. Check CXR for catheter position. PA rupture from over-wedging or balloon inflation in a distal branch is a potentially fatal complication of PACs.
Reference: RCOA – 2023 – Haemodynamic monitoring: PA catheter complications