ICD Magnet Effect — FRCA Final MCQ
Instant feedback + full explanation. One question, done properly.
Educational content. Not a substitute for clinical judgement or local policy.
Reveal the answer and explanation
Correct answer: E — Has no effect on the ICD
Placing a magnet over an ICD suspends tachyarrhythmia detection – the device will NOT deliver anti-tachycardia pacing or defibrillation shocks while the magnet is in place. This prevents inappropriate shocks triggered by diathermy EMI. Pacing function is usually UNAFFECTED (unlike in a pacemaker, where a magnet converts to asynchronous pacing). On magnet REMOVAL, tachyarrhythmia detection is restored (in most manufacturers – some older devices may remain disabled after magnet removal). External defibrillation pads must be placed on the patient before surgery in case real VF/VT occurs.
Reference: Association of Anaesthetists – 2023 – CIEDs and anaesthesia; RCOA – 2023 – Equipment safety