Pacemaker Magnet Response VVI — FRCA Final MCQ
Instant feedback + full explanation. One question, done properly.
Educational content. Not a substitute for clinical judgement or local policy.
Reveal the answer and explanation
Correct answer: B — VOO mode (asynchronous pacing at a fixed rate)
Placing a magnet over a pacemaker activates the 'magnet response' – this converts most pacemakers to an asynchronous mode (ignoring sensed signals). For VVI pacemakers: magnet converts to VOO mode (ventricular pacing at a fixed rate, typically 65-85 bpm depending on manufacturer, with no sensing). For DDD: converts to DOO. This ensures continuous pacing output regardless of electromagnetic interference from diathermy. The magnet should be removed after diathermy use to restore normal sensing function. Note: magnet response in ICDs is different – it disables tachycardia detection/therapy but does not affect pacing.
Reference: Association of Anaesthetists – 2023 – CIEDs perioperative management; RCOA – 2023 – Equipment