CRRT Circuit Pressure Alarms — FRCA Final MCQ
Instant feedback + full explanation. One question, done properly.
Educational content. Not a substitute for clinical judgement or local policy.
Reveal the answer and explanation
Correct answer: E — Catheter malposition with tip against the vessel wall
High venous return pressure alarms on a CRRT machine most commonly indicate kinking or malposition of the return (venous) lumen of the vascath, often with the tip abutting the vessel wall. Other causes include: catheter clot, fibrin sleeve, inadequate catheter length, or kinking at the insertion site. Management: reposition the patient, check the catheter for kinks, flush the lumens, and consider swapping the arterial and venous ports. If problems persist, the catheter may need repositioning or replacement under fluoroscopy.
Reference: ICS – 2023 – CRRT guidelines; RCOA – 2023 – ICM vascular access