A 55-year-old man with a cardiac pacemaker requires routine periodontal debridement. Which dental device should be avoided because of its potential to cause electromagnetic interference?

Educational content. Not a substitute for clinical judgement or local policy.

Reveal the answer and explanation

Correct answer: EMagnetostrictive ultrasonic scaler

**A. Magnetostrictive ultrasonic scaler** is the best answer. This type of scaler has historically been associated with electromagnetic interference with pacemakers and should be avoided where an alternative is available, particularly a hand scaler or piezoelectric unit used cautiously. The decisive feature is the **magnetostrictive** mechanism, rather than ultrasonic scaling as a whole. A high-speed air turbine is pneumatically powered, and a rubber dam clamp and mirror are non-electrical instruments. An LED composite curing light is not contraindicated: a prospective human study found no clinically significant interference with pacing or sensing from commonly used dental electrical devices, including battery-operated curing lights. Similarly, contemporary in-vivo evidence does not justify treating all ultrasonic scalers as absolutely contraindicated; the question is made discriminating by specifying the magnetostrictive type.

Reference: Elayi CS et al. “Interference between dental electrical devices and pacemakers or defibrillators: results from a prospective clinical study.” Journal of the American Dental Association, 2015. https://pubmed.ncbi.nlm.nih.gov/25637210/